Applications :
In chemical reactions and the pre-treatment of specimen, some substances are very sensitive to oxygen and water and cannot be carried out under normal circumstances. Although the chemical reactions of such substances can be carried out in a vacuum chamber, it cannot be operated, which makes the pre-treatment of the chemical reaction samples of such substances very difficult. The use of a vacuum glove box enables this type of materials to operate, react and test freely in the absence of oxygen and water, so it is widely used in chemistry, chemical, biology, biochemistry, especially in the preparation and research of certain catalysts and metallic organics (such as MO sources)
Equipment dimension:
Inner Chamber size: L1200*D750*H900mm Ext Chamber size: L2000*D805*H1880m |
Technical Parameters:
4.1 Structure | Single Room, 2 pcs gloves. |
4.2 Chamber material | 3mm SUS304 stainless steel |
4.3 Transition cabinet size | Ф390×L600mm Chamber right side with SUS304 stainless steel, Cover material: aluminium alloy Drawer can be moved inside chamber Max vacuum degree<100Pa |
4.4 Tools transition cabinet | Ф150×L360mm Chamber right side and material with SUS304 stainless steel, Drawer can be moved inside chamber Max vacuum degree<100Pa |
4.5 Equipment rack | 1 set, Height 800~900mm can be adjustable. |
4.6 Sample rack | 1 set,Pendant or ceiling type( double layers) |
4.7 Observe panel | 12mm |
4.8 Gloves | 2 pcs, 0.4mm butyl |
4.9 Gloves port | 2 pcs with PPS. |
4.10 Power socket | Inside of chamber, 220V,10A |
4.11 Light | 1 pcs of light |
4.12 Backup interface | KF40 Total 3 pcs, SUS304 stainless steel, on the left side or specify position. |
4.13 HEPA filter | 2 pcs, accuracy 0.3μm, inside of chamber to adsorbed dust |
4.14 Vacuum gauge | 2 pcs,Measure range: -0.1~0MPa |
4.15 Pressure gauge | 1 pcs to measuring the pressure inside chamber, Measure range: 0~4000Pa |
4.16 Pressure sensor | -2500 ~ 2500Pa (relative pressure) |
4.17 Foot controller | 1 set, double pedal to control the air supply and exhaust. |
4.18 Gas purification system | 1 set, single purification, |
4.19 Control system | 1 set, 7”Color touch screen |
4.20 Vacuum pump | 1 set, sucking rate: 8.4m3/h |
4.21 Water probe | 1 pcs,0~1000ppm accuracy:±2%FS Display accuracy: 0.1ppm |
4.22 Oxygen transmitter | 1 pcs,0~1000ppm accuracy:±2%FS Display accuracy: 0.1ppm |
4.23 Organic solvent filter | 17L, |
4.24 Main Parameters | The standard under 1 atmosphere, 25℃ ±5℃ Water content: ≤1ppm Oxygen content:≤1ppm Leakage content:≤ 0.05vol%/h |
4.25 Power requirement | 1. 220V,16A (Glove box power source) 2. 220V,10A( Socket inside box power source) |